Infrared Moisture Analyzer FD-800

Product Catalogs Available
Videos Available

Full spec model mounted with
dual temperature sensor

  • Measurement, Analysis, and Inspection:Moisture meters

Overview of Products, Technologies, and Services

Flagship model of infrared moisture analyzer
Minimum moisture content display 0.01%, weight display 1mg
Dual temperature control even monitors on-tray temperatures
Significantly shortened measurement time compared with our other models

Product Features

Feature 1 Dual temperature sensor
In addition to environmental temperature control using a conventional thermistor (T1), a radiation thermometer (T2) was employed to directly measure a new sample temperature, thus allowing high level temperature control that makes the maximum use of the merits of both types.
Feature 2 A new type of auto tare mechanism.
Since the measurement is performed automatically while taking the zero-point calibration of the balance, scale drift is compensated even over long measurement times, allowing high precision measurements.
Feature 3 Equipped with a mid-wavelength infrared quartz heater
The quartz heater (maximum energy wavelength of 2.6 μm) has excellent drying efficiency for a wide range of samples and is considered to be less prone to differences due to sample color, making it ideal for drying. In addition, the average life is 20,000 to 30,000 hours, which is 5 to 10 times longer than that of conventional infrared or halogen lamps.
Feature 4 High precision measurement.
Since the sample temperature can be measured, it is possible to set optimal drying conditions that match the characteristics of the sample, avoid burning or quality changes, and enable a more accurate moisture measurement.